BHTA member B. Braun has issued a Field Safety Notice concerning known software-related conditions that may interrupt infusion therapy in Infusomat Space, Infusomat Space P, and Perfusor Space devices operating with software versions prior to N09.
The notice provides safety information and confirms that software versions prior to N09 are no longer supported for software maintenance. It is not a product recall and does not require device removal or a mandatory software update.
Under specific operating circumstances, known software-related conditions may cause device alarms and interrupt infusion delivery, unrelated to user handling. When an alarm occurs, infusion delivery stops and audible and optical signals alert the user.
B. Braun states that the alarms for Infusomat Space and Perfusor Space devices are identical to those described in the instructions for use and should be handled in the same way.
Temporary interruption of infusion therapy may be clinically relevant depending on the patient’s condition, medication, infusion rate, and clinical setting. In certain circumstances, prolonged interruption could lead to serious patient harm if appropriate clinical intervention is not performed.
B. Braun states that available post-market and risk assessment information indicates that serious outcomes remain unlikely, and that the overall residual risk related to software conditions remains acceptable and is continually monitored.
If life-saving medication is being administered, users must, in accordance with the instructions for use, ensure supplemental patient monitoring and have a second infusion pump immediately available for the relevant drug.
B. Braun has discontinued software maintenance support for versions prior to N09. These versions will no longer receive software updates, patches, or technical software support for software maintenance purposes.
General device service, complaint handling, vigilance reporting, and safety-related investigations remain available.
Software versions currently available on the market remain acceptable for continued use. B. Braun recommends implementing N09 where feasible.
The information should be shared with all relevant departments, including clinical, technical, and biomedical personnel, as well as distribution partners.
Where products have been distributed to a third party, recipients should forward a copy of the information or inform B. Braun’s contact person identified in the notice.
Recipients are asked to complete and return the feedback form to Catherine Clulow at recalls.uk@bbraun.com. For acknowledgement enquiries, she can be contacted on 0114 2259155. The form requests confirmation that the notice has been received and a copy sent to all affected customers.
Users should review and document installed software versions in accordance with their organisation’s maintenance and quality procedures, and ensure appropriate local awareness of alarm response actions.
Healthcare institutions should evaluate N09 implementation against their clinical use cases, device configurations, enabled software functions, and local risk management procedures. Users should also assess local maintenance, validation, training, and workflow requirements.
N09 has been released and distributed since July 2025 and provides corrections for known root causes, improvements, new features, and ongoing software support. Information on obtaining and implementing it is available from the responsible B. Braun Service.
For technical information, contact Andrew Bodman, Technical Customer Support Manager, at andrew.bodman@bbraun.com or on 07977 248089.
BHTA member Convatec is voluntarily recalling specific lots of ESENTA Stomahesive Paste due to the potential presence of foreign matter.
The Field Safety Notice, dated October 2026 and carrying reference TW-3258356, requires affected products to be identified, quarantined, and removed from use and distribution.
Convatec’s investigation found that controls at a third-party manufacturing facility were not fully effective in preventing contamination of its materials.
Foreign matter may affect product quality and performance. Potential adverse effects include local inflammatory response, minor infection, skin irritation, rash, mucosal irritation, or allergic reaction.
At the time of the notice, Convatec had received no reports of related adverse events. One customer complaint involving foreign matter prompted the investigation and corrective action.
The recall covers multiple product codes and lots internationally. The table below lists only the UK-specific products identified in Appendix 2, all supplied in 60g tubes.
| Product SAP code | Product ICC code | Affected UK lot | Expiry date |
| 1740002 | 183910 | 6C03980 | 1 March 2031 |
| 1740002 | 183910 | 6D00839 | 1 April 2031 |
Readers should consult the full Convatec Field Safety Notice for the global affected product list and country-specific details. Appendix 3 shows how to identify product information on the packaging.
Distributors and healthcare facilities should check inventory without undue delay and immediately quarantine affected stock, stopping its distribution, sale, transfer, and use.
Relevant personnel must be informed. Distributors should immediately identify and notify customers supplied with affected products, retain notification records, and confirm completion upon request.
The Appendix 1 response form must be returned to European.CustomerSupport@convatec.com within 30 days, even if no affected stock is held.
Convatec will provide a Certificate of Destruction where affected inventory is identified; this must be signed and returned following destruction. Organisations that purchased through another supplier should contact that supplier for disposal instructions.
Sales representatives should immediately familiarise themselves with the notice and affected lots, stop distribution and promotion of affected products, and check personal stock, consignment stock, and demonstration kits.
They must return the Appendix 1 response form to European.CustomerSupport@convatec.com within 30 days and complete and return a Certificate of Destruction.
Representatives should support customers with stock identification where requested and escalate complaints, quality concerns, potential adverse events, and customer questions through established Convatec processes.
Users and patients should stop using affected products immediately, set them aside, and contact their healthcare provider, supplier, pharmacy, or Convatec representative. Users should keep the product details and lot number until they receive further instructions.
Anyone experiencing unexpected symptoms after using affected product, including irritation, rash, inflammation, or signs of infection, should contact their healthcare professional promptly.
Healthcare professionals should identify, quarantine, and dispose of affected products, discontinue their use, inform relevant staff, and report adverse events or complaints to Convatec and applicable regulatory authorities as required.
Convatec has received no reports indicating that patient intervention is required solely because of this issue. Professionals should use clinical judgement when assessing potentially exposed patients.
The notice must be shared with relevant internal recipients and organisations supplied with potentially affected products.
Appendix 4 of the FSN lists country-specific contacts for customer services and replacement product enquiries. In the UK, contact uk.customerservice@convatec.com or 01244 284882.
Questions about the field action can be sent to fsca-id@convatec.com.
Plans to reduce duplication in medical device regulation, improve product traceability, and provide targeted support for SMEs are among the commitments most relevant to BHTA members in the MHRA’s newly published 2030 strategy.
The ‘MHRA Strategy 2030’ sets out how the Medicines and Healthcare products Regulatory Agency intends to modernise regulation while strengthening patient safety. For members manufacturing, importing, or distributing medical devices, key points include changes to routes to market, proposed data requirements, and access to regulatory support.
MHRA Chief Executive Lawrence Tallon said: “Our 2030 Strategy sets out how, working with our partners across health and life sciences, we will translate scientific and technological innovation into better health outcomes for people across all four nations of the UK.”
Several measures remain proposals or depend on parliamentary approval; the strategy does not, by itself, introduce these changes.

Following its 2026 consultation, the MHRA is considering indefinite recognition in Great Britain of devices compliant with the EU Medical Device Regulation and In Vitro Diagnostic Medical Devices Regulation.
For members sourcing devices from the EU or supplying both British and EU markets, this could reduce duplicated regulatory work and support continuity of supply. However, indefinite recognition remains under consideration; the strategy does not confirm that all CE-marked devices will be recognised indefinitely.
The MHRA also plans to explore mutual recognition of medical devices with the US Food and Drug Administration and join the Medical Device Single Audit Program. This programme allows a single audit of a manufacturer’s quality management system to address the requirements of participating regulatory authorities, potentially reducing repeated audits for businesses operating internationally.
The strategy places greater emphasis on monitoring device performance and managing risks once products are in use.
Subject to the parliamentary process, proposed regulations would require unique device identifiers (UDIs), or an equivalent system, for most devices. The strategy specifies exceptions for non-implantable custom-made devices and non-implantable investigational devices.
The purpose is to enable precise identification and traceability throughout a device’s lifecycle. For manufacturers and suppliers, consistent identifiers could make it easier to link product records with reported safety concerns and identify affected devices. Depending on the final requirements, members may need to adapt how identifiers are recorded and shared across the supply chain.
The MHRA also plans to upgrade or replace systems for device registration, reporting, and public access to information. These changes are intended to improve data quality and transparency, giving businesses and healthcare organisations access to more reliable device information.
Greater use of health data, AI-enabled analysis, and targeted market surveillance will support earlier identification of potential risks. Improvements to Yellow Card reporting, including a more conversational interface from 2027, are intended to make reporting concerns easier for patients and healthcare professionals.
The MHRA plans targeted vouchers to help UK start-ups access premium regulatory services at lower cost. Its medical device commitments also include limited fee vouchers for UK SMEs, subject to approval.
For eligible BHTA members, this could make specialist advice more accessible. The strategy does not specify voucher values, discount levels, or detailed eligibility criteria.
The agency also intends to broaden scientific advice, including lighter-touch options for focused questions, and introduce formal Scientific Opinions providing a documented MHRA view on specific regulatory issues. These services are intended to help businesses make development decisions with greater confidence.
Subject to parliamentary approval, proposed new powers would allow the MHRA to undertake regulatory assessments directly. The strategy notes that these assessments can currently only be undertaken by Approved Bodies.
For manufacturers, this could provide an additional assessment route through the MHRA. However, the strategy does not establish that all devices could use this route or that Approved Bodies would cease to be needed. It highlights innovative technologies, including AI-enabled devices, advanced diagnostics, robotics, and novel implants.
The strategy also carries forward plans for an Early Access Service for Devices, first outlined in 2025. This would provide time-limited, conditional access for innovative devices addressing unmet clinical needs ahead of full regulatory assessment. For members developing eligible products, it would offer an opportunity to gather evidence in real-world settings while progressing towards full regulatory approval.
The MHRA plans closer coordination with national partners, including the National Institute for Health and Care Excellence (NICE), the NHS, and the Office for Life Sciences (OLS), to better connect regulation, research, assessment, and adoption.
Members developing digital or AI-enabled medical devices should note plans for an MHRA Centre for AI and Digital Health, intended to provide clearer and more consistent regulatory expectations.
The strategy envisages continuous monitoring of real-world performance, staged authorisation, and expanded Predetermined Change Control Plans. These plans would support the management of anticipated changes to AI-enabled devices, reflecting the challenges of regulating technologies whose performance can evolve over time.
The MHRA intends to strengthen patients’ contributions to regulatory decisions and publish a refreshed Patient and Public Involvement and Engagement Plan.
For assistive technology businesses, this is relevant to how users’ experiences and priorities inform the assessment of products. Greater patient involvement could help regulatory decisions reflect the outcomes that matter to people using devices in everyday life.
Read the full strategy and the MHRA’s guidance on what it means for medtech and the life sciences industry.