Field Safety Notice issued for Etac Clean 60cm Shower Commode and heightening kits

Last Updated on 25/11/2026 by Sarah Sarsby

BHTA member Etac has issued an urgent Field Safety Notice following a small number of reports that a swivel castor may come loose on a Clean Shower Commode fitted with heightening spacers. To reduce the risk of injury, Etac is voluntarily offering a set of four replacement screws for each potentially affected commode and a set of four for each potentially affected heightening kit.

What is the issue and which products are affected?

Etac traced the issue to a batch of screws used during assembly that did not meet its specification. If a swivel castor comes loose while the commode is in use, there is a risk of injury.

The notice applies only to Clean Shower Commodes fitted with heightening spacers. The spacers are fitted to the Clean 60cm Shower Commode before delivery or supplied separately as a heightening kit for fitting after purchase. The table on page two of the notice lists the product codes and affected distribution ranges for the lagoon green and white versions of the commode and kit.

What should users and healthcare facilities do?

Etac advises that the commode can continue to be used safely if all four screws are present and secure. If any screw is missing or loose, the commode must be quarantined immediately. Page two of the notice shows the four screw locations and how to replace the screws.

Etac recommends replacing all four screws even if they appear secure. In that case, replacement can take place at a convenient opportunity, such as during servicing. The replacement screws are a few millimetres longer than the original screws, which provides a stronger thread lock.

Etac asks recipients to share the notice with all relevant personnel and any organisations to which the devices may have been supplied.

Further information

Anyone with questions about the notice should contact their local Etac sales representative or distributor.